Code-card insertion or removal was too slow. Repeat Code-card insertion without
interruption.
12. Flashing result
If temperature gradient is too high, SensoLite Nova still keeps on operating, but
the displayed result will be flashing, indicating the abnormal operational circumstances.
Be cautious with these test results, they might be inaccurate.
13. Instrument shows Lo constantly
Code setting is not correct. Check and set the proper code value.
Insufficient amount of blood was applied on the strip. Repeat test with a bigger drop of blood
14. Instrument shows Lo constantly
Code setting is not correct. Check and set the proper code value.
15. Battery symbol appears on display
Batteries are low. Replace batteries.
If error messages cannot be eliminated by applying the recommendations above, please
contact your local distributor for advice.Thank you for your confidence in purchasing
our product.