Blood Glucose Meter
Instructions
for Use
In vitro Diagnostic Medical Device
IVD
SensoLite Nova
SensoLite Nova PLus
0197
H
u
m
a
n
V
o
i
c
e
D
a
t
a
T
e
l
l
i
n
g
Seitenansicht 0
1 2 3 4 5 6 ... 58 59

Inhaltsverzeichnis

Seite 1 - SensoLite Nova

Blood Glucose MeterInstructionsfor UseIn vitro Diagnostic Medical DeviceIVDSensoLite NovaSensoLite Nova PLus0197HumanVoiceDataTelling

Seite 2

9SensoLite Nova Plasma calibrated SensoLite Nova meter displays the result of blood glucose measurementas it had been obtained from the plasma reading

Seite 3

10SensoLite Nova IMPORTANT PRECAUTIONSAlways operate SensoLite Nova blood glucose meter according to the instructionsdescribed hereafter. If the equip

Seite 4

11SensoLite Nova TECHNICAL DATAPower supply: 2 pcs of 3V CR2032 type batteries Battery lifetime: approx. 1 year or 1000 measurementsCalibrated for: Se

Seite 5

12SensoLite Nova PARTS OF THE METER 1. OK button 6. Code-card slot2. Up button 7. Battery holder3. Down button 8. IR data output interface4. Display 9

Seite 6

13SensoLite Nova ABOUT TEST STRIPSensoLite Nova is designed ONLY for in vitro use with SensoLite Nova Test strip.No other strips will give you an accu

Seite 7

14SensoLite Nova SensoLite Nova Test strips are sensitive to heat and mechanical damage – do not usestrips if you suspect such danger may occur. Alway

Seite 8

15SensoLite Nova Strip insertion and useThe strip must be inserted into meter's strip holder with the black side facing upwards andwith the squar

Seite 9

16SensoLite Nova BEFORE STARTING MEASUREMENTReadings might be effected by: Skin contaminants: Always wash and dry your hands thoroughly before taking

Seite 10

17SensoLite Nova BLOOD SAMPLINGIt is recommended to use the lancing device and disposable lancets packed together withthe meter to prick your finger.

Seite 11

18SensoLite Nova STRIP-CODE SETTINGYour SensoLite Nova meter and SensoLite Nova Test strip form together an accuratemeasuring system. To assure that t

Seite 12

1SensoLite Nova CONTENTSIMPORTANT – READ THIS BEFORE USING THE EQUIPMENT 3INTRODUCTION 4PACKING LIST 5SENSOLITE NOVA PLUS METER EXPLAINED 6UNIQUE FEAT

Seite 13

19SensoLite Nova  If your Code-card is broken or missing, please follow instructions for ”Setting strip-code manually” in this chapter!Setting strip-

Seite 14

20SensoLite Nova 2. Press the Down button to enter the ”code setting” function.An animation displayed on the screen will request the Code-cardto be in

Seite 15

21SensoLite Nova 4. The meter returns automatically to the test screen where the recently set code value isdisplayed, showing that it has been stored

Seite 16

22SensoLite Nova Setting strip-code manuallyWhen the Code-card is missing or it is damaged, you may want to set the strip code manu-ally. The range of

Seite 17

23SensoLite Nova Code value can be adjusted digit by digit. After entering this menu function modify the valueof the first digit with the Up and Down

Seite 18

24SensoLite Nova MEASURING WITH THE METERBefore you start the test make sure you have everything ready you might need: meter, teststrip, finger pricki

Seite 19

25SensoLite Nova entrance of the strip holder. If the strip is in its right place thescreen changes: instead of the flashing arrow a flashing blood dr

Seite 20

26SensoLite Nova 3. Flashing blood drop icon disappears from the screen after theblood droplet reaches the strip. Flashing clock icon shows that theme

Seite 21

27SensoLite Nova 6. After completing the measurement, test strip can beremoved from the meter. Hold the meter vertically and turn the ejector wheel up

Seite 22

28SensoLite Nova Warning! High or low results which are incorrect may have serious medical consequences. If your blood glucose is unusually high (gre

Seite 23

2SensoLite Nova MEASURING WITH THE METER 24MEMORY 29Storing data 29Recalling data from memory 30Average calculation 31OVERVIEW OF FUNCTION MENUS 32SET

Seite 24

29SensoLite Nova MEMORYSensoLite Nova is able to store up to 500 measurements with related times and dates.Storing dataSensoLite Nova meter automatica

Seite 25

30SensoLite Nova Recalling data from memoryStored data can be retrieved from memory by selecting MEMORYmode. Press the Up button after switching the m

Seite 26

31SensoLite Nova Average calculationSensoLite Nova is able to calculate the average glucose level of the last 7-14-28 daystests' results. Calcula

Seite 27

32SensoLite Nova OVERVIEW OF FUNCTION MENUSBy keeping the OK button pressed after switching on, meter will offer alternatively six func-tional modes:

Seite 28

33SensoLite Nova Switch meter on with the OK button and keep it pressed. Select the time setting menu by releasing the OK button when “SEt” time appea

Seite 29

34SensoLite Nova When time is set press the OK button again to go on to date setting. In the upper rightcorner first month then day will be displayed.

Seite 30

35SensoLite Nova Switch the meter on with the OK button and keep it pressed until “PC” mode appears. Release the button.Target LiteLink adapter's

Seite 31

36SensoLite Nova SELECTING MEASUREMENT UNIT(mmol/l – mg/dl)SensoLite Nova meter is able to display measured results in two measurement systemunits: mm

Seite 32

37SensoLite Nova After entering this menu only the recently active unit will appear on the screen. Switch between the two possible units (mmol/l and m

Seite 33

38SensoLite Nova DELETING DATA STORED IN MEMORYAll data in memory can be deleted in dEL mode. After deletiondefault data setting will be restored (set

Seite 34

3SensoLite Nova IMPORTANT – READ THIS BEFORE USINGTHE EQUIPMENTBefore testing your blood with SensoLite Nova or SensoLite Nova Plus Blood GlucoseMonit

Seite 35

39SensoLite Nova SWITCHING TO SPEAKING FUNCTIONSensoLite Nova Plus meter can speak in two differentlanguages by default: English and German. To switch

Seite 36

40SensoLite Nova BATTERY REPLACEMENTSensoLite Nova meter is powered by 2 pcs of round cell batteries, type CR2032. Theircapacity is sufficient for abo

Seite 37

41SensoLite Nova If 10 test measurements are done after the first appearance of thebattery icon, the error message "E-6" will be displayed o

Seite 38

42SensoLite Nova Important! If the instrument does not work after battery replacement: replace the batteries again, press and hold the OK button down

Seite 39

43SensoLite Nova Warning! Protect the inside of the meter from water!  Never immerse the meter or hold it under running water! Avoid disassembling

Seite 40

44SensoLite Nova Checking with Check-stripPress OK button and keep it pressed until “CAL” appears on thescreen, then release it.Strip icon will be dis

Seite 41

45SensoLite Nova If the displayed value is within the given range – that is indicatedon the label affixed to the Check-strip holding pocket –, themete

Seite 42

46SensoLite Nova Checking with CareSens control solutionPress OK button and keep it pressed until “CAL” appears on the screen, then release it.Strip

Seite 43

47SensoLite Nova Shake control solution bottle well before use! Drop CareSenscontrol solution to the reagent area of test strip, which is at thetip o

Seite 44

48SensoLite Nova If the result of self test is not in the given value range: perform self test again. check if the code value displayed on LCD is th

Seite 45

4SensoLite Nova INTRODUCTIONThank you for choosing SensoLite Nova blood glucose monitoring system. It has been care-fully designed to enable you to ea

Seite 46

49SensoLite Nova Important! Do not continue test if an error message is displayed!This chapter lists error messages, explains their means and possible

Seite 47

50SensoLite Nova 3. E-O error message If this error message remains on screen after turning meter on, the meter has got da- maged. Please remove the

Seite 48

51SensoLite Nova  Meter is damaged. Check the meter with Check-strip, see “SensoLite Nova system checking” chapter.7. E-6 error message Batteries a

Seite 49

52SensoLite Nova 10. E-9 error message External temperature is too high or low (out of the range 15-40 °C). Wait at least 30 minutes before using the

Seite 50

53SensoLite Nova  Code-card insertion or removal was too slow. Repeat Code-card insertion without interruption.12. Flashing result If temperature g

Seite 51

54SensoLite Nova MANUFACTURER'S WARRANTY77 Elektronika Kft. warrants SensoLite Nova blood glucose meter against defects in materialsand workmansh

Seite 52

55SensoLite Nova SYMBOLS0197 98/79/EEC IVD directive, SensoLite Nova meter0120 93/42/EEC MDD directive, lancetIn vitro diagnosticumInformation insideW

Seite 53

56SensoLite Nova Manufacturer:77 ELEKTRONIKA Kft.1116 Budapest, Fehérvári út 98. HUNGARYTel: + 36 1 206 - 1480Fax: + 36 1 206 - 1481E-mail: sales@e77.

Seite 54

57SensoLite Nova

Seite 55

77 Elektronika Kft.Fehérvári út 98. H-1116 BudapestTel.: +36 1 206-1480Fax: +36 1 206-1481E-mail: [email protected]. e77.hu2.0 version 2006. 02.

Seite 56

5SensoLite Nova PACKING LISTContent of SensoLite Nova kit SensoLite Nova blood glucose meter (D47-8001-x) Instruction For Use (D47-9201-x) Instruct

Seite 57

6SensoLite Nova SENSOLITE NOVA METER EXPLAINEDSensoLite Nova indicates blood glucose concentration by checking the reaction betweenchemical reagents a

Seite 58

7SensoLite Nova UNIQUE FEATURE OF SENSOLITE NOVA PLUSThis feature is a great help for those people who may have problems with reading displayedcharact

Seite 59

8SensoLite Nova FOR USERS HAVING PLASMA EQUIVALENTCALIBRATED METERThis section refers only to users who have plasma equivalent calibrated meter. If yo